奧林巴斯纖維胃鏡維修常見故障判斷

來源:http://m.thefunfoodfactory.com/ 發布時間:2024-03-10 瀏覽量:0

鉗道管漏水:是由于使用活檢鉗等附件不當造成鉗道破損,屬 C 級修理 。

Leakage of forceps: It is caused by improper use of biopsy forceps and other accessories, resulting in damage to the forceps and classified as a C-level repair.

導管束斷:使用年限較長,彎曲管不良造成,屬 C+ 導光束級修理 。

Tube bundle breakage: caused by long service life and poor bending of the tube, it belongs to C+beam level repair.

20200426035244690

彎曲角度不到位:鋼絲前端斷:使用年限長,磨損造成, C 級修理 。

Inadequate bending angle: Broken front end of steel wire: Long service life, caused by wear and tear, C-level repair.

彎曲角度不到位:鋼絲后端脫焊和使用一定時間后鋼絲變長,D 級修理,

Inadequate bending angle: The rear end of the steel wire is detached and the steel wire becomes longer after a certain period of use, D-level repair,

(在彎曲好的情況)。

(In a well bent condition).

彎曲橡皮漏水:碰傷造成比較多, D 級修理更換專用橡皮即可 。

Bending rubber leakage: It is caused by more scratches. D-level repair can replace the specialized rubber.

噴嘴堵塞:清洗不干凈, D 級修理更換噴嘴 。

Nozzle blockage: cleaning not clean, D-level repair to replace nozzle.

導像束黑點多或龜裂:纖維鏡使用年限長或保養不當造成,A 級修理 。

Multiple black spots or cracks in the imaging beam: caused by long service life or improper maintenance of the fiber mirror, A-level repair.

電子內窺鏡鏡頭模糊:具體故障需經漏水測試,烘干,拆鏡檢查仔細判別才能定如何修理(嚴重時要做 A 級修理)。

The lens of the electronic endoscope is blurry: specific faults need to be carefully judged through water leakage testing, drying, and disassembly inspection before determining how to repair (in severe cases, A-level repair should be performed).

電子鏡打角度時有網狀出現:使用年限長 CCD 根部導線斷造成,那一定要做 A 級大修。

When adjusting the angle of the electronic mirror, there is a mesh like appearance: it is caused by a broken wire at the root of the CCD due to its long service life. Therefore, it is necessary to undergo an A-level major repair.

s
新聞資訊
聯系我們
  • 公司公司:匠仁醫療設備有限公司
  • 匠仁醫療山東濟南維修部
  • 聯系人:樊經理
  • 電話:13153199508
  • 地址:山東省濟南市槐蔭區美里東路3000號德邁國際信息產業園6號樓101-2室
  • 匠仁醫療湖南長沙維修部
  • 聯系人:李經理
  • 電話:13873135765
  • 地址:湖南省長沙市雨花區勞動東路820號恒大綠洲14棟2409室
新聞資訊

樊經理:13153199508     李經理:13873135765

公司地址:山東省濟南市槐蔭區美里東路3000號德邁國際信息產業園6號樓101-2室  湖南省長沙市雨花區勞動東路820號恒大綠洲14棟2409室

     

魯公網安備 37010402441281號